Use of Thermoluminescence Dosimetry for QA in High-Dose-Rate Skin Surface Brachytherapy with Custom-Flap Applicator
Sensors(2023)
摘要
Surface brachytherapy (BT) lacks standard quality assurance (QA) protocols. Commercially available treatment planning systems (TPSs) are based on a dose calculation formalism that assumes the patient is made of water, resulting in potential deviations between planned and delivered doses. Here, a method for treatment plan verification for skin surface BT is reported. Chips of thermoluminescent dosimeters (TLDs) were used for dose point measurements. High-dose-rate treatments were simulated and delivered through a custom-flap applicator provided with four fixed catheters to guide the Iridium-192 (Ir-192) source by way of a remote afterloading system. A flat water-equivalent phantom was used to simulate patient skin. Elekta TPS Oncentra Brachy was used for planning. TLDs were calibrated to Ir-192 through an indirect method of linear interpolation between calibration factors (CFs) measured for 250 kV X-rays, Cesium-137, and Cobalt-60. Subsequently, plans were designed and delivered to test the reproducibility of the irradiation set-up and to make comparisons between planned and delivered dose. The obtained CF for Ir-192 was (4.96 ± 0.25) μC/Gy. Deviations between measured and TPS calculated doses for multi-catheter treatment configuration ranged from −8.4% to 13.3% with an average of 0.6%. TLDs could be included in clinical practice for QA in skin BT with a customized flap applicator.
更多查看译文
关键词
surface brachytherapy,thermoluminescent dosimeter,custom applicator,quality assurance,radiation dosimetry,in vivo dosimetry
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
数据免责声明
页面数据均来自互联网公开来源、合作出版商和通过AI技术自动分析结果,我们不对页面数据的有效性、准确性、正确性、可靠性、完整性和及时性做出任何承诺和保证。若有疑问,可以通过电子邮件方式联系我们:report@aminer.cn