Reliability of IMU-Derived Gait Parameters in Foot Drop Patients

CARING IS SHARING-EXPLOITING THE VALUE IN DATA FOR HEALTH AND INNOVATION-PROCEEDINGS OF MIE 2023(2023)

引用 0|浏览20
摘要
Foot drop is a deficit in foot dorsiflexion causing difficulties in walking. Passive ankle-foot orthoses are external devices used to support the drop foot improving gait functions. Foot drop deficits and therapeutic effects of AFO can be highlighted using gait analysis. This study reports values of the major spatiotemporal gait parameters assessed using wearable inertial sensors on a group of 25 subjects suffering from unilateral foot drop. Collected data were used to assess the test-retest reliability by means of Intraclass Correlation Coefficient and Minimum Detectable Change. Excellent test-retest reliability was found for all the parameters in all walking conditions. The analysis of Minimum Detectable Change identified the gait phases duration and the cadence as the most appropriate parameters to detect changes or improvements in subject gait after rehabilitation or specific treatment.
更多
查看译文
关键词
Foot Drop,Ankle-Foot Orthosis,Gait Analysis,Inertial Measurement Unit,Intraclass Correlation Coefficient