Reproducibility and Accuracy of a Pocket-Size Ultrasound Device in Assessing Liver Steatosis.
DIGESTIVE AND LIVER DISEASE(2024)
摘要
Background: This diagnostic prospective study compared the feasibility and diagnostic accuracy of Pocketsize Ultrasound Devices (PUDs) against standard ultrasound (US) in detecting liver steatosis using the controlled attenuation parameter (CAP) and liver biopsy as reference standards. Materials and Methods: Consecutive patients with chronic liver diseases were assessed for the presence of steatosis using PUD and US. A CAP cut-off value > 275 dB/m was applied to establish >S1. A 26-patient subgroup underwent liver biopsy. PUD reproducibility was evaluated using Cohen's k statistic. Diagnostic accuracy of PUD and US was given as Sensibility (Sn), Specificity (Sp), Positive and Negative Predictive Values (PPV, NPV), positive and negative Likelihood Ratio (LR + , LR-). Results: 81 consecutive patients (69% males) with multiple etiologies were enroled. PUD inter-observer agreement was good (k 0.77, 95%CI 0.62-0.93). PUD and US identified >S1 according to CAP values respectively with Sn 0.87, Sp 0.61, PPV 0.49, NPV 0.91, LR+ 2.04, LR- 0.07, AUROC 0.74 and Sn 0.96, Sp 0.54, PPV 0.47, NPV 0.97, LR+ 2.10, LR- 0.07, AUROC 0.75. Conclusions: PUD shows good reproducibility and diagnostic accuracy in ruling liver steatosis out, representing a useful point-of-care tool to avail of hepatologists interested in excluding NAFLD, but with basic US skills. (c) 2023 Editrice Gastroenterologica Italiana S.r.l. Published by Elsevier Ltd. All rights reserved.
更多查看译文
关键词
Steatosis,Non-alcoholic fatty liver disease,Pocket size ultrasound devices,Ultrasound,Controlled -attenuation parameter,Vibration -controlled transient elastography,Liver biopsy
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
数据免责声明
页面数据均来自互联网公开来源、合作出版商和通过AI技术自动分析结果,我们不对页面数据的有效性、准确性、正确性、可靠性、完整性和及时性做出任何承诺和保证。若有疑问,可以通过电子邮件方式联系我们:report@aminer.cn