Cerebral Embolic Protection in Transcatheter Aortic Valve Implantation Using the Sentinel Cerebral Protection System: A Systematic Review and Meta-Analysis
Cardiology and Therapy(2024)
摘要
Transcatheter aortic valve implantation (TAVI) plays a vital role in patients with symptomatic aortic stenosis. Despite the mortality benefit of TAVI, embolic stroke remains a feared complication. As a result, transcatheter cerebral embolic protection (TCEP) devices have been developed to reduce this risk. Given the ongoing debate of TCEP in TAVI, we performed a systematic review and meta-analysis of all randomized controlled trials to date to identify outcomes of periprocedural stroke using the Sentinel™ cerebral protection system (CPS). MEDLINE, Cochrane, and Scopus databases were utilized from inception until 12/2023. PRISMA criteria was utilized. Keywords included “cerebral embolic protection”, “sentinel cerebral protection system”, “transcatheter aortic valve implantation”, and “transcatheter aortic valve replacement”. Primary outcome was periprocedural stroke. Secondary outcomes included periprocedural disabling and non-disabling stroke, all-cause mortality, transient ischemic attack, delirium, acute kidney injury, vascular complications, bleeding, and pacemaker implantation. Risk ratios (RR) were measured via Mantel–Haenszel method with fixed analysis. Heterogeneity was assessed via chi-squared and Higgin’s I2 test. Four trials with 3528 patients were assessed. SAPIEN 3 was the most common bioprosthetic valve used. The average age was 79.4 years with 41.9
更多查看译文
关键词
Cerebral embolic protection,Sentinel cerebral protection system,Transcatheter aortic valve implantation,Transcatheter aortic valve replacement,Periprocedural stroke
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
数据免责声明
页面数据均来自互联网公开来源、合作出版商和通过AI技术自动分析结果,我们不对页面数据的有效性、准确性、正确性、可靠性、完整性和及时性做出任何承诺和保证。若有疑问,可以通过电子邮件方式联系我们:report@aminer.cn